Home page News page : Last Updated 27-Nov-2007 Project Description page Data page Sensors page Contacts page Calendar page Join page Links to related sites Partner Area
 

In April 2008 the ANIMATE network Observatories became part of the EuroSITES European Ocean Observatory Network.
Data shown here and subsequent data are available through the EuroSITES website www.eurosites.info
The ANIMATE site will not be updated further.

Fluorimeter Data Summary

 

Click here to read the report "Calibration of chlorophyll-fluorescence during ANIMATE" by Susan Hartman

Summary of Fluorimeter sensor Deployments

Click on the highlighted calibration figures in the cells of the table to view the complete calibration document.

Deployment

CIS

ESTOC

PAP

1         Cruise

      Serial No

 

P293/2 Aug 2002

H2020341

Calibrated Chl(ug/l)=95.106*fl676+0.0507

R2=0.8069

Data and sensor lost June 2003

M53/1b April 2002

H2020340

No initial calibration. Different sensor mount to PAP and CIS. Mooring broke May 2002. 250 days data downloaded. Some –ve values from day 13 and some noise (not seen in lab investigations)

D266 Sept 2002

H2020343

Calibrated Chl(ug/l)=311.63*fl676+0.004

R2=0.96

 -ve on calibration dip, noise and slight hysteresis Heavy deposits on emitter/detector so data unreliable

2         Cruise

      Serial No

 

P302 Aug 2003

H2020342

Calibrated on PAP2

Chl (ug/l)=59.431 *fl676+0.0102

R2=0.8751

No long term data

P296 April 2003

H2020340

Calibrated  Chl(ug/l)=699.45*fl676+0.0052

R2=0.88

Chl analysed ICCM. HS2

Good data

P300/1 July 2003

H2020343

Redeployed (Cu oxide scraped off, gap reset) Calibration for unit H2020342 on P300/1, but not for H2020343.10 months good data downloaded, some –ve values.

3         Cruise

      Serial No

 

Not deployed

P305 Oct 2003

H2020340

Calibrated  Chl(ug/l)=1009.9*fl676+0.0709

R2=0.6051

Chl analysed ICCM. HS2

Good data.

P306 Nov 2003

H2020343

Calibrated Chl(ug/l)=314.2246 * fl676 - 0.0072

R2=0.9855

5 months good data

4         Cruise

      Serial No

 

CD161 Sept 2004

Wet Labs  FLNTUSB-179

2 stage calibration:

 Flu=(0.0127 *(signal-55);

 Chl(ug/l) = (0.4259*Flu)+ 0.1434

 R2=0.9663

P310 April 2004

H2020340

Battery pack not working

CD158 June 2004

H2020343

Calibrated  Chl(ug/l)=(280.92 * fl676) + 0.109

R2=0.941

5

November 2005

WETLabs FLNTUSB-270

Unit ran from 17-Sep-2005 to 19-Nov-2005 then failed, found to be flooded on recovery.

P319 Dec 2004

H2020342

Calibrated Chl(ug/l)=393.01*fl676+0.0143

R2=0.9682

Chl analysed ICCM. HS2

D295t July 2005

FLNTUSB 237

6

D309 Aug 2006

FLNTUSB 268

Calibrated at PAP D295t

Calibrated Chl (ug/l) = 0.241*sensor chl +0.0827

R2= 0.8344

 

Corrected calibration at CIS

Chl (ug/l) = (0.7531 * Corrected Chl_signal) + -0.0211

R2= 0.9845

P320 Mar 2005

H2020340

Calibrated Chl (ug/l)=(290.94*fl676)+0.1005

R2=0.4245

HS2

D306 July 2006

Sensor package not deployed due to fishing risk

7

MSM05/4 July 2007

FLNTUSB 515

P330 Nov 2005

CE07/16

FLNTUSB 270

Mooring failed after one month, but sensors recovered

8

 

P333 March 2006 FLNTUSB 238 deployed but did not record any data

9

 

P344-2 Nov 20065

WetLAB 238

FLNTUSB_Chl (ug/l) = 0.0123*(chlorophyll_signal-53)

Chl (ug/l)= 0.2981 * FLNTUSB_Chl + 0.2123

R2=0.92

10

 

MSM05/4 March 2007
FLTNTUSB 179 failed on calibration dip

11

 

Unknown

12

 

P365/1 Recovered Sensors and removed DOLAN

 

 

Data may be found at the ftp site ftp://ftp.noc.soton.ac.uk/pub/animate/, or on the CD, in the animate_data directory

      Back to Top of page